Assay Method Information

Assay Name:  Folate Receptor SPR Assay
Description:  Table 16: Folate receptor SPR binding assays were conducted on a Biacore 8K instrument (Cytiva) using biotinylated folate receptor beta (Acro Biosystems). Briefly, folate receptor was immobilized at -550 RU on a streptavidin chip (Cytiva) in 10 mM HEPES pH 7.5, 150 mM NaCl, 0.05% Tween 20. Multi-cycle kinetic studies were conducted with 8 compound concentrations using 2-fold dilutions in 10 mM HEPES pH 7.5, 150 mM NaCl, 0.05 % Tween 20, 2% DMSO at 25° C. A flow rate or 75 μl/mmn with an association time of 120 s and a dissociation time of 600 s were used. Surfaces were regenerated with 4 M MgCl2 at 30 μl/min for 30 s. Data were fit to a 1:1 model to determine the reported binding parameters.
Affinity data for this assay
 