Assay Method Information | |
| Binding Assay |
Description: | All the FP based experiments were performed in MICROFLUOR 2 Black, "U" Bottom, 96-well Microtiter Plates (ThermoSci.) and FP was measured as mP units in a microplate reader (Tecan Ultra) with excitation at 485 nm and emission at 530 nm. To dilutions of WDR5Δ23 (2.2-0 uM) in 100 ul assay buffer (0.1M Phosphate, 25 mM KCl, 0.01% Triton, pH 6.5) 20 ul of a fixed concentration of the tracer in the assay buffer was added, followed by an addition of 5 ul DMSO to give 125 ul of total volume. Each assay had two controls: blank (without protein and tracer) and tracer only (without protein). The plates were incubated on a shaker at room temperature to reach equilibrium and mP values were measured at the 3 hour time point. |
Affinity data for this assay | |
---|---|
If you find an error in this entry please send us an E-mail |